Troponin I Rapid Test
The RapidFor™ Troponin I rapid test is used for the qualitative detection of cardiac troponin I (cTnI) in whole blood, serum or plasma.
82,50 €
Quantity | Price | Discount |
---|---|---|
5-20 | 78,38 € | 5% |
21-50 | 74,25 € | 10% |
51+ | 66,00 € | 20% |
RapidFor™ Troponin I Rapid Test
The RapidFor™ Troponin I Rapid Test is a rapid chromatographic immunoassay for the qualitative detection of cardiac troponin I (cTnI) in whole blood, serum or plasma. It is used as an aid in the diagnosis of acute myocardial infarction (AMI). The test is easy to perform, provides results within 10-15 minutes and is ideal for use in emergency departments, doctors’ surgeries and clinical laboratories.
Troponin I is a heart-specific protein and a highly sensitive marker for myocardial damage. It is detectable in the blood 4-6 hours after an infarction, reaches its peak within 12-24 hours and remains elevated for up to 14 days. This makes cTnI ideal for the diagnosis and monitoring of heart attacks.
Clinical performance data
The clinical performance of the RapidFor™ Troponin I rapid test was analysed by comparison with an established ELISA test procedure. A total of 598 patient samples were tested.
- Sensitivity: 98,68 %
- Specificity: 99,32 %
- Accuracy: 99,00 %
The minimum detection limit (LOD) of the test was determined to be 0.5 ng/mL. This allows the detection of even low concentrations of troponin I in the blood.
No cross-reactions were detected with:
- Skeletal muscle troponin I (10 μg/mL)
- Cardiac troponin T (2000 ng/mL)
- Cardiac myosin (20 μg/mL)